BA200-Fully Auotomated Biochemistry Analyser
Click Here To See the Product Brochure
| Parameter |
Typical Specification |
| General Performance |
- Throughput: 200 test/h (without electrolytes)
- Throughput ISE module: 120 test/h (3 channels), 160 test/h (4 channels)
- Principles of analysis: Colorimetry, turbidimetry, ISE method: Na+, K+,CI (L+ as optional)
|
| Reactions rotor |
- Minimum reaction volume: 180 µL
- Minimum reaction volume: 440 µL
- Number of wells: 120
- Well material: UV methacrylate
- Type of incubation: 5 min (fixed)
- Temperature reaction rotor: 37° C
- Accuracy of temperature: ± 0.2° C
- Temperature stability: ± 0.1° C
- Mixers: 1
|
| Cuvette washing system |
- Number of tips of washing system: 7
- Number of tips of washing solution: 2
- Rinsed with water: 3
- Dried tips: 2
- Wash volume per tip: 711 µL
- Consumption of washing solution: 1.42 mL/cycle (=7.11 µL cws)
|
| Environmental Requirements |
- Ambient temperature: 10°C to 35°C, 10°C to 30°C (With ISE module)
- Relative humidity: <85% without condensation.
|
| Dimensions and weight |
- Dimensions (width, depth and height): 1077 mm x 690 mm x 680 mm
- Weight: 166 Kg
|
| Electrical Requirements |
- Mains voltage: 115 V to 230 V
- Mains frequency: 50 Hz or 60 Hz
- Electric power: Max. 500 VA
- Average power: 260 VA
|
| Fluidic Requirements |
- Water inlet: External tank or mains water supply
- Type of water: Purified type II (NCCLS)
- Water consumption: <9 L/h
- High concentration waste tank: 2.4L
- Washing solution tank: 2.4L
|